Promag CON-485 Series Technical Information Seite 9

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 40
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 8
Proline Promag 50W, 53W
Endress+Hauser 9
Cable specifications
remote version
Coil cable:
2 x 0.75 mm
2
PVC cable with common, braided copper shield (Ø approx. 7 mm)
Conductor resistance: 37 /km
Capacitance: core/core, shield grounded: 120 pF/m
Permanent operating temperature: –20…+80 °C
Cable cross-section: max. 2.5 mm
2
Signal cable:
3 x 0.38 mm
2
PVC cable with common, braided copper shield (Ø approx. 7 mm) and individually
shielded cores.
With Empty Pipe Detection (EPD): 4 x 0.38 mm
2
PVC cable with common, braided copper shield
(Ø approx. 7 mm) and individually shielded cores.
Conductor resistance: 50
/km
Capacitance: core/shield: 420 pF/m
Permanent operating temperature: –20…+80 °C
Cable cross-section: max. 2.5 mm
2
A0003194
a = signal cable, b = coil current cable (cross-section: max. 2.5 mm
2
)
1 = core, 2 = core insulation, 3 = core shield, 4 = core jacket, 5 = core strengthening,
6 = cable shield, 7 = outer jacket
Optionally, Endress+Hauser also supplies reinforced connecting cables with an additional, metal strenghtening
braid. We recommend such cables for the following cases:
Cables laid underground
Danger of rodent attack
Device used with ingress protection IP 68
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with EN 61010, the EMC
requirements of EN 61326/A1, and NAMUR Recommendation NE 21.
Caution!
Grounding is by means of the ground terminals provided for the purpose inside the connection housing.
Keep the stripped and twisted lengths of cable shield to the terminals as short as possible.