Proline Promag 50W, 53W
Endress+Hauser 3
Input
Measured variable Flow rate (proportional to induced voltage)
Measuring range Typically v = 0.01...10 m/s with the specified measuring accuracy
Operable flow range Over 1000 : 1
Input signal Status input (auxiliary input):
U = 3…30 V DC, R
i
= 5 k, galvanically isolated.
Configurable for: totalizer(s) reset, measured value suppression, error-message reset.
Status input (auxiliary input) with PROFIBUS DP and MODBUS RS485:
U = 3…30 V DC, R
i
= 3 k, galvanically isolated
Switching level: 3...30 V DC, independent of polarity
Configurable for: totalizer(s) reset, measured value suppression, error-message reset,
batching start/stop (optional), batch totalizer reset (optional)
Current input (for Promag 53 only):
Active/passive selectable, galvanically isolated, full scale value selectable,
resolution: 3 µA, temperature coefficient: typ. 0.005% o.r./°C (o.r. = of reading)
active: 4...20 mA, R
i
150 , U
out
= 24 V DC, short-circuit-proof
passive: 0/4...20 mA, R
i
150 , U
max
= 30 V DC
Output
Output signal Promag 50
Current output:
active/passive selectable, galvanically isolated, time constant selectable (0.01...100 s),
full scale value selectable, temperature coefficient: typ. 0.005% o.r./°C (o.r. = of reading),
resolution: 0.5 µA
active: 0/4…20 mA, R
L
< 700 (HART: R
L
250 )
passive: 4…20 mA, operating voltage V
S
18...30 V DC, R
i
150
Pulse/frequency output:
passive, open collector, 30 V DC, 250 mA, galvanically isolated.
Frequency output: full scale frequency 2...1000 Hz (f
max
= 1250 Hz), on/off ratio 1:1,
pulse width max. 10 s.
Pulse output: pulse value and pulse polarity selectable, max. pulse width configurable
(0.5...2000 ms)
PROFIBUS DP interface:
Transmission technology (Physical Layer): RS485 in accordance with ANSI/TIA/EIA-485-A: 1998,
galvanically isolated
Profile version 3.0
Data transmission rate: 9.6 kBaud...12 MBaud
Automatic data transmission rate recognition
Function blocks: 1 x analog input, 3 x totalizer
Output data: volume flow, totalizer
Input data: positive zero return (ON/OFF), totalizer control, value for local display
Cyclic data transmission compatible with previous model “Promag 33”
Bus address adjustable via miniature switches or local display (optional) at the measuring device