Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning
Endress+Hauser 27
5.6 Set-up using ATA flash card
Save the device settings on the PC onto the ATA flash card using the operating and readout soft-
ware. This set-up file can be adopted into the device, if released, under “Extras - ATA flash - Load
ATA flash set-up”.
!
Note!
In order to be able to use this function, the device must already be registered in the PC database.
"
Caution!
If the set-up ATA flash card is not removed, the measurement data will begin to be saved after
approx. 5 minutes. The set-up data is still kept. Please change the ATA flash card if the measurement
data is not meant to be saved onto this card.
!
Note!
The functions described above are also valid for the use of Compact Flash cards.
5.7 List of operating parameters
5.7.1 Basic settings
Settings that are generally valid, e.g. date, time, etc.
Set-up -> Basic settings
Device designation Use this to describe, for example, where the device is installed (important if you are using several devices). 20 characters.
! Note!
Is also saved onto the ATA flash card. In the PC, graphics/tables are provided with this text (important, for example, if
you have several devices in operation).
The device designation is also available for the export, e.g. in a spreadsheet program.
Current date Format: DD.MM.YY
Current time Format: hh:mm, 24 h display
Summertime changeover Function of summertime / normal time changeover.
“Automatic”: changeover as per valid EC directives
“Manual”: Set the changeover times in the next operating positions
“Off”: no time changeover
NT/ST Region In Europe, the summertime / normal time changeover takes place at different times to that in the USA.
Select the region in which the device is installed.
Date NT -> ST Day in spring on which the changeover from normal time to summertime takes place.
Format: DD.MM.YY